Q: Does OneDose™ affect the cost of treatment?
A: If the patient is eligible for Medicare and Medicaid, the cost of dosimetry should be covered. Prescribed properly, dosimetry should be covered. However, there are national coverage limits; if a patient requests additional dosimetry beyond the national coverage limits, it would then affect the cost of treatment.

Q: Is OneDose™ covered by health insurance?
A: Yes; it has a national reimbursement code. We recommend submitting a claim for OneDose™ using CPT code 77331 for Special Dosimetry (e.g., TLD, microdosimetry) (specify) only when prescribed by the treating physician. Please look under the reimbursement heading for more detailed information.

Q: Can the OneDose™ patch be reused?
A: No. It is used one time to ensure an accurate read.

Q: How will using OneDose™ benefit me?
A: It provides the radiation therapy team a more reliable method of monitoring dose, including a system for permanently recoding the readings.

Specifications
• Temperature for dosimeter and reader storing/operating: 10º - 40º C
• OneDose Energy Range: Co60, 6 - 18 MV photon
                                                             6 - 16 MeV electron
• OneDosePlus Energy Range:     6 - 18 MV photon
• Dose Range: 0 - 500 cGy
• Accuracy: <30 cGy +/- 2 cGy
                     30 - 500 cGy, +/-5.0% (2 sigma), under full build-up conditions for OneDose, no added build-up for OneDosePlus
• OneDose Dosimeter Dimensions: 6 mm wide, 33 mm long, and <1 mm thick.
• The detectors are pre-calibrated to Co60 under full buildup conditions (refer to Calibration Certificate).